Plasma/Low Hb System

analyzer
  • Accurate evaluation of hemolysis

  • Dedicated support and service

  • Easy to use


Specifications
Principle Modified azide-methemoglobin reaction; dual wavelengths (570 nm and 880 nm) for compensation of turbidity
Calibration Factory calibrated against the ICSH reference method; needs no further calibration
Sample Material Plasma, serum and aqueous solutions, or stored erythrocyte suspensions
Measurement Range 0 – 30.0 g/L (0 – 3.0 g/dL, 0 – 3000 mg/dL, 0 – 1.9 mmol/L)
Results Including data entry, within one minute for normal glucose levels
Sample Volume ~20 µL
Dimensions 160 × 210 × 90 mm (6.29 × 8.26 × 3.54 inches)
Weight 1 kg (2.0 pounds) with batteries installed
Storage Temp. Analyzer: 0 – 50 °C (32 – 122 °F) Microcuvettes: unopened 15 – 30 °C (59 – 86 °F); three-month open vial stability
Operating Temp. 15 – 30 °C (59 – 86 °F)
Power AC Adapter or 4 AA batteries
Quality Control Two levels of liquid controls

Advantages

analyzer

Plasma/Low Hb System

analyzer

Confidence answers

Precise factory calibration against the ICSH reference method and patented microcuvette technology with excellent lot-to-lot reproducibility

Easy access to lab-quality accuracy

The results are ready within one minute, testing anywhere you need (portable system)

3  SIMPLE STEPS

Fill

microcuvette.

Put

microcuvette into analyzer.

See

the results are ready.