About
analyzer
Hb 201 DM System
analyzer
-
Complete data management solution
-
Improved patient safety with features like QC lock-out and built-in barcode scanning for patient ID
Specifications
Principle | Modified azidemethemoglobin reaction; dual wavelengths (570 nm and 880 nm) for compensation of turbidity | |
Calibration | Factory calibrated against the ICSH reference method; needs no further calibration | |
Sample Material | Capillary, venous or arterial whole blood | |
Measurement Range | 0.5- 25.6 g/dL (5- 256 g/L, 0.3- 15.9 mmol/L) | |
Results | 15- 60 seconds | |
Sample Volume | -10 μL | |
Dimensions | Analyzer: 170 x 93 x 50 mm (6.70 x 3.66 x 1.97 inches) Docking Station: 135 x 206 x 61 mm (5.30 x 8.10 x 2.40 inches) |
|
Weight | Analyzer: 350 g (0.77 pounds) with batteries installed Docking Station: 566 g (1.24 pounds) |
|
Storage Temp. | Analyzer: 0-50 °C (32 – 122 °F) Microcuvettes: unopened 15 – 30 °C (59 – 86 °F); three-month open vial stability |
|
Operating Temp. | 18-30 °C (64-86 °F) | |
Power | Internal rechargeable batteries or powered with AC adapter direct or in docking station | |
Interface | Printer and HemoCue® Basic Connect including optional barcode scanner. Data transfer using Bluetooth® technology is possible via HemoCue® BT Connect (accessory) | |
Quality Control | Built-in self-test, optional liquid controls |
4 SIMPLE STEPS
Fill
microcuvette.
Put
microcuvette into analyzer.
See
the results are ready.
Link
seamlessly with your network.
Data Management Features
Data Entry | Touch-screen display
Built-in barcode scanner |
|
Interface | Interface with existing software solutions using POCT1-A (CIC standard). Docking solution enables connection of up to 5 analyzers | |
Configurable Functions | Operator ID, Patient ID, Lot ID,
Patient test comments Critical value alerts STAT testing |
|
Result Storage | 4,000 patient/STAT tests,
500 QC tests, 500 analyzer logs, 200 patient IDs |
|
Quality Control | QC lock-out | |
Data Management | QC scheduling Linearity and proficiency testing |
Advantages
Hb 201 DM System
Improve Administrative Workflow
Seamlessly interface with existing network using POCT1-A. Ensure regulatory compliance and simplify billing
Safeguard Patient Testing
Link patient information and comments with test result. Reduce errors with automatic result transfer
Easy access to lab-grade accuracy
Precise factory calibration against the ICSH reference method and microcuvette technology with excellent lot-to-lot reproducibility
Hb 201 DM